| Model Number |
3303-90151 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Failure to Eject (4010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an aleutian tlif ii straight inserter and inner shaft were unable to be removed from an implant intra-operatively.The tip of the inserter inner shaft subsequently fractured off inside the implant.The implant and fractured inner shaft tip were left inside the patient.This report captures the inserter.
|
| |
|
Event or Problem Description
|
|
It was reported that an aleutian tlif ii straight inserter and inner shaft were unable to be removed from an implant intra-operatively.The tip of the inserter inner shaft subsequently fractured off inside the implant.The implant and fractured inner shaft tip were left inside the patient.This report captures the inserter.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon visual inspection it was noted that there was no damage to the aleutian tlif ii straight inserter, it was the inner shaft that sustained damage.Report #3004774118-2022-00203 was filed for the inner shaft.
|
| |
|