• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION COUPLER; DEVICE, ANASTOMOTIC, MICROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION COUPLER; DEVICE, ANASTOMOTIC, MICROVASCULAR Back to Search Results
Catalog Number 511100150010
Medical Device Problem Code Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the ring of a 1.5mm coupler was not able to fully attach and failed to complete anastomosis.This issue was identified during use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.The white plastic holder, 2.5mm jaw assembly, both 2.5mm rings were free floating in the inner coupler tray.Visual inspection was performed which observed that the rings potentially had an excess amount of tissue everted onto the pins that would not allow the rings to fully mate as intended to complete the anastomosis.The reported condition was verified.The cause of the condition could not be determined.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER
Common Device Name
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave. w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14465756
Report Number1416980-2022-02579
Device Sequence Number17324072
Product Code MVR
UDI-Device Identifier00844735002652
UDI-Public(01)00844735002652
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K861985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number511100150010
Device Lot NumberSP21E131534655
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/13/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-