SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 50MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 74122550
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 04/16/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, left bhr-tha was performed on (b)(6) 2007 due to arthritis, the patient experienced pain and significant high cobalt and chromium ions.This adverse event was treated with a revision surgery on (b)(6) 2021, in which trunionosis and small pseudotumor formation were observed.The patient's condition is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that a left bhr-tha revision surgery was performed due to pain and significant high cobalt and chromium ions, trunionosis and small pseudotumor formation were observed intraoperatively.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the devices concerned was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for hemi head and the modular sleeve.However, as the devices are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.The hemi head and the modular sleeve involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure mode.A risk management review was performed.The alleged failure mode and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.Although it was reported the patient had high cobalt and chromium ions, neither the levels nor the lab reports were provided for review.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported clinical reactions cannot be confirmed.It cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant or implant failure.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 50MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14465914
Report Number3005975929-2022-00308
Device Sequence Number14601565
Product Code KWL
UDI-Device Identifier00885556071205
UDI-Public885556071205
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74122550
Device Lot Number11353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-