Model Number VPR-GW-FLEX18
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
When attempting to remove the diamondback 360 peripheral orbital atherectomy device (oad), the viperwire advance guide wire fractured.The oad was removed, but the fragment remained in vivo with a filter deployed over the fragment.The filter detached from the wire while attempting to snare the fragment.The fragment and filter were successfully retrieved, and nothing remained in vivo.
 
Additional Manufacturer Narrative
Return of the device for analysis is anticipated.If the device is received for analysis, a supplemental report will be submitted when the device analysis is completed.Medwatch report (b)(4) was submitted for this event.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE)
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
kim mcclure
1225 old hwy 8 nw
st. paul, MN 55112
6519006741
MDR Report Key14465934
Report Number3004742232-2022-00123
Device Sequence Number4569381
Product Code MCW
UDI-Device Identifier10852528005725
UDI-Public(01)10852528005725(17)231031(10)403317-2
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberVPR-GW-FLEX18
Device Catalogue Number7-10041-04
Device Lot Number403317-2
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Initial Report FDA Received Date05/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-