• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TIBIAL PLATE, PROXIMAL MEDIAL AXSOS 3 FOR RIGHT TIBIA 4 HOLE / L71MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TIBIAL PLATE, PROXIMAL MEDIAL AXSOS 3 FOR RIGHT TIBIA 4 HOLE / L71MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 627734
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the proximal lateral tibia plate and screws were broken.Due to an acute infection, the broken implants, a medial plate and additional screws, were removed and resolved with temporary external fixation placement.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since no evidences that the infection was product related were provided, also there is no allegation against the device in question.The device inspection revealed the following: the visual inspection has shown that the plate has some slight wear marks at the surfaces and the threads, these marks can be traced back to normal use during insertion/extraction.Otherwise is the device in a good condition, no damage that could have contributed to the reported infection or the breakage plate and the screws could be detected.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.No product related issue could be detected.Based on the available information it is not possible to define the root cause of the reported infection, it can be confirmed that the device in question was supplied in unsterile condition to the customer.In this relation following statement from the instructions for use can be pointed out: ¿warning in the event of contamination, or expiration of shelf life or in the case of products supplied non-sterile, the product must be subjected to an appropriate cleaning process and sterilized by means of a validated sterilization procedure before use, unless specified otherwise in the product labeling or respective product technical guides.¿ if more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that the proximal lateral tibia plate and screws were broken.Due to an acute infection, the broken implants, a medial plate and additional screws, were removed and resolved with temporary external fixation placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL PLATE, PROXIMAL MEDIAL AXSOS 3 FOR RIGHT TIBIA 4 HOLE / L71MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14465945
Report Number0008031020-2022-00244
Device Sequence Number4507077
Product Code HRS
UDI-Device Identifier07613327093889
UDI-Public07613327093889
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K200398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number627734
Device Catalogue Number627734
Device Lot NumberL20330
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexMale
-
-