Model Number PBD-203-0707
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was sent to olympus for evaluation.Evaluation of the device found the product and labeling did not match.The product label, barcode label, and triple sticker showed pbd-203-0707, and the pack printing and product body showed pbd-203-0704.Based on the results of the legal manufacturer's investigation, it is likely the product listed on the carton did not match the product body due to multiple products being placed in one sterilization basket and sterilized at the same time.The pbd-203-0704 was then packed as pbd-203-0707, because pbd-203-0704 was mixed in with the pbd-203-0707.However, a definitive root cause could not be determined.A corrective and preventive action (capa) has been opened to further investigate this issue.Olympus will continue to monitor field performance for this device.G3: investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.
 
Event or Problem Description
The customer reported that when inspecting the biliary stent before use during an unspecified therapeutic procedure, the contents of the box were different from the outer box.The intended procedure was completed using another similar device.There was no patient or user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information provided by the legal manufacturer.A review of the device history record (dhr) for the subject lot indicated no anomaly with the event-related items.The record includes the following: process inspection sheet, quality inspection sheet.Although there were no anomalies identified in the dhr, the capa (previously mentioned in the initial report) was opened because of the actual discrepancy between the box and its contents due to manufacturing causes.A search was performed to determine whether there were other complaints received for mispackaging of the products, and no other complaints were received for this issue.Activities have been taken to investigate this event and prevent reoccurrence of this event in the future.
 
Additional Manufacturer Narrative
Correction to g3 of the initial medwatch.The aware date should be (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE BILIARY STENT V
Common Device Name
SINGLE USE BILIARY STENT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14466142
Report Number8010047-2022-08685
Device Sequence Number4516005
Product Code FGE
UDI-Device Identifier04953170186370
UDI-Public04953170186370
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K933200
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPBD-203-0707
Device Lot Number0XKA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer06/23/2022
11/08/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date06/30/2022
12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-