• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS,HIP,SEMICONSTRAINED,METALCERAMICCERAMICMETAL,CEMENTEDORUNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS,HIP,SEMICONSTRAINED,METALCERAMICCERAMICMETAL,CEMENTEDORUNCEMENTED Back to Search Results
Model Number 71335552
Medical Device Problem Codes Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hyperplasia (1906); Metal Related Pathology (4530)
Date of Event 08/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case-2022-00103691-1.
 
Event or Problem Description
It was reported that, after right r3-tha was performed on (b)(6) 2010 due to right severe osteoarthritis, the patient experienced degenerative changes and mild synovial hyperplasia.This adverse event was treated with a revisions surgery on (b)(6) 2021.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.The pictures provided were reviewed and could not confirm the metallosis.The clinical/medical investigation concluded that, with the information provided the clinical root cause of the reported metallosis cannot confirmed.However, it is unknown if the retroverted acetabular component led to accelerated wear and the reported metallosis.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for hip systems provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.It is unknown if the retroverted acetabular component led to accelerated wear and the reported metallosis.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include damaged product, implant corrosion or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that, after right r3-tha was performed on (b)(6) 2010 due to right severe osteoarthritis, the patient experienced degenerative changes and mild synovial hyperplasia.When the acetabular component was retroverted, metallosis was confirmed.This adverse event was treated with a revisions surgery on (b)(6) 2021.Patient's current health status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 52MM
Common Device Name
PROSTHESIS,HIP,SEMICONSTRAINED,METALCERAMICCERAMICMETAL,CEMENTEDORUNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14466161
Report Number1020279-2022-02483
Device Sequence Number11904584
Product Code MRA
UDI-Device Identifier03596010598240
UDI-Public03596010598240
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/06/2020
Device Model Number71335552
Device Catalogue Number71335552
Device Lot Number10HM05219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-