• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) VERITY ADX XL; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5256
Medical Device Problem Code Failure to Disconnect (2541)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented for a routine pacemaker change.During implant, the physician was unable to unscrew the connector pin of the ventricular lead from the pacemaker header.Several attempts were made but there was no rotation of the screw on the pacemaker header.There was no complaint on the lead only on the pacemaker.The lead was capped.Both the pacemaker and lead were replaced successfully.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of the setscrew could not be untightened to remove the lead was confirmed.Analysis revealed that calcified blood was filling the ventricle setscrew inset.The calcified blood was preventing the wrench from engaging the setscrew inset to untighten the setscrew.Wrenches cannot penetrate the calcified blood.The wrench will just strip the portions of the inset it comes in contact with.When the calcified blood was removed from the inset a wrench could be fully inserted into the inset and untightened the setscrew.The lead could then be removed.The blood came through a hole punched through the septum by the torque wrench at time of implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITY ADX XL
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14466602
Report Number2017865-2022-10660
Device Sequence Number14624137
Product Code DXY
UDI-Device Identifier05414734006811
UDI-Public05414734006811
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2010
Device Model Number5256
Device Catalogue Number5256
Device Lot Number2918010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ISOFLEX OPTIM LEAD