The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was reported that a patient with a 19mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 4 years, 3 months due to restricted leaflet motion, severe aortic stenosis and mild aortic regurgitation.The patient presented with cardiogenic shock.There was concern for av thrombus and the patient was started on eliquis for one month prior to the tavr.The tavr was performed with a 23mm 9750tfx transcatheter valve successfully.The patient tolerated the procedure well without complications.
|
|
H10: additional manufacturer narrative: updated sections d4 (expiration date), h4, h6 (type of investigation, investigation findings, investigation conclusions).Leaflet immobility or leaflet restriction occurring over time, and not due to extrinsic physical interference, is a form of structural valve deterioration, which can result in significant regurgitation and/or stenosis.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Degeneration-related structural deterioration is most commonly related to patient factors and is not usually an indication of a device malfunction.The root cause of this event was determined to be due to patient related factors.This event was impacted by the progression of the patient's underlying valvular disease pathology with structural valve deterioration including ckd stage 3 and hld.The subject device is not available for evaluation, as it remains implanted in the patient.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|