EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 3000TFX
Medical Device Problem Codes Fluid/Blood Leak (1250); Gradient Increase (1270); Difficult to Open or Close (2921)
Health Effect - Clinical Codes Cardiogenic Shock (2262); Insufficient Information (4580)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 19mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 4 years, 3 months due to aortic stenosis.The tavr was performed with a 23mm 9750tfx transcatheter valve successfully.The patient did well.
 
Event or Problem Description
It was reported that a patient with a 19mm 3000tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 4 years, 3 months due to restricted leaflet motion, severe aortic stenosis and mild aortic regurgitation.The patient presented with cardiogenic shock.There was concern for av thrombus and the patient was started on eliquis for one month prior to the tavr.The tavr was performed with a 23mm 9750tfx transcatheter valve successfully.The patient tolerated the procedure well without complications.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: updated sections d4 (expiration date), h4, h6 (type of investigation, investigation findings, investigation conclusions).Leaflet immobility or leaflet restriction occurring over time, and not due to extrinsic physical interference, is a form of structural valve deterioration, which can result in significant regurgitation and/or stenosis.Tissue degeneration-related structural deterioration, either calcific or non-calcific, are common chronic failure modes for this type of bioprosthetic heart valve.Operational mechanical stress and biological factors are generally believed to be the major contributors to the non-calcific bioprosthetic tissue degeneration.Structural valve deterioration (svd) can, and typically does, lead to chronic central leaks over a period of time.Svd is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Degeneration-related structural deterioration is most commonly related to patient factors and is not usually an indication of a device malfunction.The root cause of this event was determined to be due to patient related factors.This event was impacted by the progression of the patient's underlying valvular disease pathology with structural valve deterioration including ckd stage 3 and hld.The subject device is not available for evaluation, as it remains implanted in the patient.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
1 edwards way
mailstop ant 6.1
irvine, CA 92614
9492502731
MDR Report Key14466622
Report Number2015691-2022-05817
Device Sequence Number4686212
Product Code DYE
UDI-Device Identifier00690103163636
UDI-Public(01)00690103163636(17)200204
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P860057/S02
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/04/2020
Device Model Number3000TFX
Device Catalogue Number3000TFX19MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer05/27/2022
07/18/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date06/21/2022
07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/05/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age68 YR
Patient SexFemale
Patient Weight50 KG
-
-