Model Number 154006
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the plastic cylinder at the bottom of the drain bag released with little to no pressure.It was also stated that normally the lever holds it in but this drain bag did not have a de-tent.As per the follow up via phone on (b)(6) 2022, it was reported that the customer did not have the defective drain bag lot number, but they have the sample to return.It was stated that no medical interventions or concerns occurred concerning this event.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device met relevant specifications.The product was used for patient treatment.The product had not caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used drainage bag with inlet tube and sample port connector.Visual inspection of the sample noted drainage bag was filled with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the bag outlet device open and closed as intended.No root cause could be found because the reported event was unconfirmed.A device history record review was not required as the investigation was unconfirmed.The investigation is concluded, and no additional action is required at this time.As the reported event is unconfirmed a labeling review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the plastic cylinder at the bottom of the drain bag released with little to no pressure.It was also stated that normally the lever holds it in but this drain bag did not have a de-tent.As per the follow up via phone on 05may2022, it was reported that the customer did not have the defective drain bag lot number, but they have the sample to return.It was stated that no medical interventions or concerns occurred concerning this event.Per additional information received via sample form on 25may2022, stated that it did not stay in closed position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URINARY DRAINAGE BAG WITH ANTI-REFLUX CHAMBER
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14466650
Report Number1018233-2022-03993
Device Sequence Number4504709
Product Code KNX
UDI-Device Identifier00801741029615
UDI-Public(01)00801741029615
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number154006
Device Catalogue Number154006
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-