• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II XR BASEPLATE SZ 4 RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II XR BASEPLATE SZ 4 RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Model Number 74022294
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Damage to Ligament(s) (1952)
Date of Event 04/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed on (b)(6) 2019, the patient tore their anterior cruciate ligament (acl).The adverse event was addressed via revision surgery on (b)(6) 2022, in which the jrny ii xr baseplate sz 4 rt was explanted along with the insert.The condition of the patient is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a revision was performed approximately 2.5 years post implantation due to ¿patient tore their anterior cruciate ligament (acl)¿.Reportedly, a jrny xr baseplate was replaced/revised along with the insert.The requested medical documentation has not been received as of the date of this medical investigation and the current patient status is unknown.Without the requested clinical documentation, the root cause of the reported acl tear could not be further assessed.The patient impact beyond the reported injury and subsequent revision of the baseplate and insert could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed in warnings and precautions that the patient should be warned of surgical risks, and made aware of possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, and/or patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II XR BASEPLATE SZ 4 RT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14466689
Report Number1020279-2022-02487
Device Sequence Number4516021
Product Code JWH
UDI-Device Identifier00885556615423
UDI-Public00885556615423
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74022294
Device Catalogue Number74022294
Device Lot Number18HSL0017A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date06/27/2022
Date Device Manufactured05/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-