| Model Number |
CD2311-36Q |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented to the hospital for a generator change procedure on (b)(6) 2022.During interrogation of the implantable cardioverter defibrillator (icd), it was noted that there was overestimation of battery longevity.No programming changes were made to the icd.The patient was in stable condition.
|
| |
|
Event or Problem Description
|
|
New information received notes the implantable cardioverter defibrillator (icd) was explanted and replaced on (b)(6) 2022.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b1, b2, b5, d6b, d9, h1, h3, h6.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2022-22377.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of overestimation was confirmed.The device was at elective replacement indicator (eri) upon receipt.Longevity analysis found the battery depletion to be normal.The device functioned normally during analysis.No anomalies were found.The longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
|
| |
|