• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VIVACIT-E DM BEARING 28X40MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VIVACIT-E DM BEARING 28X40MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Hyporesponsive to stimuli (2683)
Date of Event 04/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Report source: source: (b)(6).Concomitant medical products: catalog number:802202802 lot number: 64816019 brand name: femoral head 12/14.Multiple reports were submitted along with this report 0002648920-2022-00129.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient was revised due to dislocation and disassociation of the bearing which was free floating in the joint causing loose of sensation in the hip approximately 20 days post implantation.Additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: image one demonstrate superior dislocation.Image two demonstrates polyethylene wear.Image three demonstrates possible metallosis.Image three also demonstrates a small metallic focus near the greater trochanter, possibly loose body.Visual examination of the returned product identified nicks gouges and scratches were noted to the inner and outer spherical surface as well as the rim.Rim also shows deformation from head wear.No other damage was noted to the liner.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.A summary of the investigation was requested and sent to the reporter.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVACIT-E DM BEARING 28X40MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14468421
Report Number0001822565-2022-01489
Device Sequence Number4570151
Product Code LPH
UDI-Device Identifier00889024572683
UDI-Public(01)00889024572683(17)270112(10)65254480
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K190656
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110031010
Device Lot Number65254480
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-