• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. MAXFORCE MTP PLATE, LONG, 0°-0°, LEFT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. MAXFORCE MTP PLATE, LONG, 0°-0°, LEFT; PLATE, FIXATION, BONE Back to Search Results
Model Number MAXFORCE MTP PLATE, LONG, 0°-0°, LEFT
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 5/3/2022, it was reported by a sales representative via phone that an ar-9944l-0l maxforce mtp compression plate was implanted successfully in patient during a first mtp fusion procedure on (b)(6) 2021.Three months ago, due to daily walking, the plate broke.On (b)(6) 2022, patient underwent revision surgery and had the plate removed.Surgeon implanted an ar-8944cl-s low profile mtp plate and completed the revision surgery without any issues.Both original and revision surgery were performed at the same facility and by the same surgeon.No further issues has been reported.
 
Additional Manufacturer Narrative
The complaint is confirmed based on the customer provided photo, which displays the broken plate.However, without return of the device for physical evaluation, the cause remains undetermined.No change in harm was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXFORCE MTP PLATE, LONG, 0°-0°, LEFT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14468565
Report Number1220246-2022-04930
Device Sequence Number17324133
Product Code HRS
UDI-Device Identifier00888867345737
UDI-Public00888867345737
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K201235
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMAXFORCE MTP PLATE, LONG, 0°-0°, LEFT
Device Catalogue NumberAR-9944L-0L
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date10/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-