According to the complaint, patient blood samples for several babies were measured on the abl90 flex plus analyzer (b)(4), reporting calcium results of 0.6 mmol/l.Reference measurements from the hospital laboratory reported calcium results of 2.3 mmol/l- 2.4 mmol/l.Based on the measurements, the results from the abl90 flex plus analyzer were reported as false low by the customer.No reports of death or serious injury.
|
Radiometer medical aps has concluded their investigation.The analyzer did not malfunction, the results are due to samplers (capillaries).It was discovered that the capillaries were high heparin and are in fact the cause of the problem, the customer was unaware that 200 iu indicated the heparin content of the capillaries.The customer is now going to check across their other sites as one of them has now reported the same issue.
|