• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a shoulder arthroscopy both anchor broke off inside the patient during insertion.The broken pieces were retrieved from patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection it was noted that on one device the suture threader was missing.The suture assembly was still assembled to the driver.On the second driver, the suture assembly was disassembled from the driver and the o-ring was not returned for investigation.On both devices it was noted that the screw was broken.On one it was broken close to the distal end of the screw.On the other it was broken on the proximal end.The method used to prepare the bone during the procedure as well as the bone quality encountered were not provided.The cause remains undetermined, however, a probable cause is attributed to improper bone preparation and/or interference between the device and hard bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14468835
Report Number1220246-2022-04936
Device Sequence Number4878423
Product Code MAI
UDI-Device Identifier00888867097919
UDI-Public00888867097919
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K203495
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberBIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Device Catalogue NumberAR-2323BCC-1
Device Lot Number14365316
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-