• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL OG HELICAL XTRSOFT COIL 2X4; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL OG HELICAL XTRSOFT COIL 2X4; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 640HX0204
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter phone: (b)(6).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.
 
Event or Problem Description
As reported by the field, during a coil embolization, a 2x4 orbit galaxy coil (640hx0204, 30571045) became unraveled/stretched in the unspecified microcatheter (mc).The physician switched to a new one to complete the surgery.There was no patient injury report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Section b5: additional information received indicated that the coil became unraveled/stretched during advancement.There was resistance encountered prior to the coil unraveling.Prior to this coil, other coils were able to be advanced through the same microcatheter.There was no blood flow restriction/reduction as a result of the event.Adequate flush was maintained through the devices.There were no procedural delays due to the event.The procedure was to the ophthalmic artery segment.The aneurysm size was 3x4.Complaint conclusion: as reported by the field, during a coil embolization, a 2x4 orbit galaxy coil (640hx0204, 30571045) became unraveled/stretched in the unspecified microcatheter (mc).The physician switched to a new one to complete the surgery.There was no patient injury report.Additional information received indicated that the coil became unraveled/stretched during advancement.There was resistance encountered prior to the coil unraveling.Prior to this coil, other coils were able to be advanced through the same microcatheter.There was no blood flow restriction/reduction as a result of the event.Adequate flush was maintained through the devices.There were no procedural delays due to the event.The procedure was to the ophthalmic artery segment.The aneurysm size was 3x4.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation (mre) was performed for the finished device 30571045 number, and no non-conformances related to the malfunction were identified.Coil unraveling/stretching is indicative of the application of excessive force, possibly in an attempt to overcome the reported resistance.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation (mre), there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG HELICAL XTRSOFT COIL 2X4
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key14469082
Report Number3008114965-2022-00357
Device Sequence Number19959445
Product Code KRD
UDI-Device Identifier10886704030713
UDI-Public10886704030713
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number640HX0204
Device Catalogue Number640HX0204
Device Lot Number30571045
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2022
Supplement Date Received by Manufacturer07/04/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNSPECIFIED MICROCATHETER
Patient SexUnknown
-
-