• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE E 10 MM THICKNESS; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE E 10 MM THICKNESS; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2022-00145.Concomitant medical products: partial tibial cemented size e left medial item# 42538000501, lot# 64201293.Partial femur cemented size 3 left medial item# 42558000301, lot# 64321307.
 
Event or Problem Description
It was reported that a patient was revised approximately ten months post implantation due to pain and decreased daily activities.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10 no product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified patient continue to experience pain and had difficulty walking.To counter this issue, patient was prescribed pain medication and using assistive device while walking.X-ray review shows no issue with patient hardware.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.Per package insert:01501264, pain is known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: a4, b4, b5, b7, g3, g6, h1, h2, and h10.
 
Event or Problem Description
It was reported that a patient was revised approximately ten months post implantation due to ongoing pain and patient comorbidities after gardening.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
Upon reassessment of this event, it was determined this product should not have been reported under this mfr number.This report should be voided, and a corrected report will be filed under mfr 0001825034-2023-00862.
 
Event or Problem Description
Upon reassessment of this event, it was determined this product should not have been reported under this mfr number.This report should be voided, and a corrected report will be filed under mfr 0001825034-2023-00862.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE E 10 MM THICKNESS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14469651
Report Number0001822565-2022-01475
Device Sequence Number17285649
Product Code HSX
UDI-Device Identifier00880304818194
UDI-Public(01)00880304818194(17)231130(10)64222252
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberN/A
Device Catalogue Number42518200510
Device Lot Number64222252
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer05/16/2022
06/09/2022
04/19/2023
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date06/10/2022
06/15/2022
04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight69 KG
-
-