Model Number UM-S20-17S
Medical Device Problem Code Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned by the customer, without the protocol observed for biohazard material, in a condition whereby the evaluation of the device could not be performed.The customer will be resending the device with the biohazard protocol observed.As such, the device is not evaluated as yet, and a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
 
Event or Problem Description
As reported for this event by the customer, during reprocessing immediately after a procedure, it was observed that there was presence of blood inside the ultrasound tip.There is no possibility that the device was used in any other procedure with the presence of blood inside the ultrasound tip.There is no harm to any patient or persons.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The subject device was returned and an evaluation completed for it where the users complaint was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that there is a perforation in the sheath where blood or body fluid may have entered the sheath.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Common Device Name
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14469713
Report Number8010047-2022-08720
Device Sequence Number20059555
Product Code ITX
UDI-Device Identifier04953170241000
UDI-Public04953170241000
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K982323
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-