Model Number SN6AT4
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
The physician reported during an intraocular lens implant procedure.A scratch was found on the lens on the posterior face, compromising the center of the optic.The lens remained in the eye since it was not compromising the visual axis.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Attached photos show the screen of a monitor.An implanted iol is observed on the monitor.The photos show what appears to be possible a linear scrape mark on the anterior optic surface.The file indicated the use of a qualified associated cartridge and viscoelastic.Per the information provided a company ii handpiece was used with the company iii d cartridge.The company ii handpiece will not physically accommodate the company iii d cartridge.This may have been reported in error.No determination can be made without physical evaluation of the complaint sample.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14469995
Report Number1119421-2022-01096
Device Sequence Number19976131
Product Code MJP
UDI-Device Identifier00380652250467
UDI-Public00380652250467
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/16/2025
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.220
Device Lot Number15122928
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II IOL DELIVERY SYSTEM, INJECTOR; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-