| Model Number |
71440145 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the gii p/s cons housing ream dome was found broken.Since this was noticed during a field inspection, no case was involved.
|
| |
|
Event or Problem Description
|
|
It was reported that the blades of the gii p/s cons housing ream dome were found deformed and dull.Since this was noticed during a field inspection, no case was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.Upon visual inspection, it was confirmed that the device was not broken, however, the reamer dome¿s blades are deformed and dull.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.H11: corrected information in b5 and h6 (medical device problem code was updated).
|
| |
|
Search Alerts/Recalls
|