• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DYONICS 4.0MM ELITE ACROMIONIZER BURR(6); SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DYONICS 4.0MM ELITE ACROMIONIZER BURR(6); SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200724
Medical Device Problem Codes Overheating of Device (1437); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy two blades made a strange noise.It occurred after about 10 seconds of rotation.The procedure was completed using an unknown device.It is unknown if there was a delay.No patient complications were reported.Overheating was reported on device with the lot number 50943904.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection revealed the device was returned outside of original packaging.Signs of overheating and melting on the sluff chamber.No other physical damage visible.A functional evaluation revealed the device functions, but makes noise due to sluff chamber becoming melted.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include rotating the device at speeds faster than intended, running the device for long periods of time without allowing it to cool, or excessive force on the device creating excess friction.No containment or corrective actions are recommended at this time.H11: corrected information in a1 (should be read in blank), b5, and h6 (health effect - impact code).
 
Event or Problem Description
It was reported that during an arthroscopy, the blade made a strange noise and overheated.It occurred after about 10 seconds of rotation.The procedure was completed using an unknown device.It is unknown if there was a delay.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 4.0MM ELITE ACROMIONIZER BURR(6)
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14470601
Report Number1219602-2022-00774
Device Sequence Number20008434
Product Code HAB
UDI-Device Identifier03596010578020
UDI-Public03596010578020
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2026
Device Model Number72200724
Device Catalogue Number72200724
Device Lot Number50943904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer12/07/2023
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date11/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-