• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-0537
Medical Device Problem Code Material Erosion (1214)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 04/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2009 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
An event regarding abnormal ion level and wear (trunnionosis) involving a accolade stem was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: "the implant sheet from the index surgery on (b)(6) 2009 confirms that the patient had a cocr v40 femoral head implanted.In the indications portion of the (b)(6) 2021 revision tha operative report the surgeon notes that pre-operative cobalt and chromium levels were elevated (levels not specified).A pre-operative mri showed a large fluid collection and erosive lesions in the greater trochanter.These findings were confirmed at time of surgery.Intra-operatively upon removal of the femoral head some "fouling" was seem on the trunnion but the trunnion was in "pretty good shape".Implants were well fixed.The trochanteric lesions were debrided and bone grafted.The patient having been previously implanted with a v40 cocr head is confirmed as is a revision surgery for trunnionosis with erosive lesions in the greater trochanter.The surgeon noted elevated serum cocr levels but did not specify the exact levels.The root cause of the trunnionosis and erosive lesions cannot be determined from this limited documentation." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.A review of the provided medical information by a clinical consultant indicated: "the implant sheet from the index surgery on (b)(6) 2009 confirms that the patient had a cocr v40 femoral head implanted.In the indications portion of the (b)(6) 2021revision tha operative report the surgeon notes that pre-operative cobalt and chromium levels were elevated (levels not specified).A pre-operative mri showed a large fluid collection and erosive lesions in the greater trochanter.These findings were confirmed at time of surgery.Intra-operatively upon removal of the femoral head some "fouling" was seem on the trunnion but the trunnion was in "pretty good shape".Implants were well fixed.The trochanteric lesions were debrided and bone grafted.The patient having been previously implanted with a v40 cocr head is confirmed as is a revision surgery for trunnionosis with erosive lesions in the greater trochanter.The surgeon noted elevated serum cocr levels but did not specify the exact levels.The root cause of the trunnionosis and erosive lesions cannot be determined from this limited documentation." the medical records provided for review were insufficient to confirm the reported events or determine root cause.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2009 and was revised on (b)(6) 2021.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14470888
Report Number0002249697-2022-00736
Device Sequence Number11942674
Product Code LPH
UDI-Device Identifier04546540510709
UDI-Public04546540510709
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2014
Device Model Number6021-0537
Device Catalogue Number6021-0537
Device Lot Number30483702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/04/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-