• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 7MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036127200
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during that during coils embolization, the subject coil was inserted and during placing, it was detached prematurely.As a result, the physician used the new coils and microcatheter to complete the procedure without clinical consequences to the patient.No further information is available.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported complaint could not be confirmed and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer was when the target xl 360 soft 7mm x 20cm was inserted, and during placing, it was detached prematurely, it was reported that the device was prepared as per the dfu, that there was no damage noted to the packaging prior to opening and the device was confirmed to be in good condition during preparation/prior to use on the patient, and that continuous flush was set up and maintained during the procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.
 
Event or Problem Description
It was reported during that during coils embolization, the subject coil was inserted and during placing, it was detached prematurely.As a result, the physician used the new coils and microcatheter to complete the procedure without clinical consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 7MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14471006
Report Number3008881809-2022-00239
Device Sequence Number17324086
Product Code HCG
UDI-Device Identifier07613327004632
UDI-Public07613327004632
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2023
Device Model NumberM0036127200
Device Catalogue NumberM0036127200
Device Lot Number223418848
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR 1018 MICROCATHETER (STRYKER).; TARGET COILS (STRYKER).
Patient SexUnknown
-
-