WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE FOR SUPRAPATELLAR; SURGICAL IMPLANT/TRIAL-IMPLANT/SIZER HOLDER, REUSABLE
|
Back to Search Results |
|
| Model Number |
03.010.440 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The lot number was unknown.Therefore, the expiration date and device manufacture date were unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported by the customer in (b)(6) that during an unknown surgery on an unknown date, while doing a suprapatellar tibial nail case, the surgeon attempted to attach the aiming arm (03.010.441) to the insertion handle (03.010.440) but it would not attach.According to the report, the insertion handle was inspected and it was found that it was bent which did not allow the two instruments to connect.It was reported that the surgeon used freehand locking technique to insert the screws.There was a surgical delay of 10 minutes to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This complaint involves one device.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary = > product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Part: : 03.010.440 lot no : 160385-101 release to warehouse date: 20 dec, 2016 manufacturing site: (b)(4).Supplier: leitner ag a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the insertion handle for suprapatellar, p/n: 03.010.440.A dimensional inspection was unable to be performed due to device design.A functional test was not conducted since the mating device was not returned for examination.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the insertion handle for suprapatellar, p/n: 03.010.440 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed manufactured dimensional inspection: n/a device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|