STRYKER TRAUMA KIEL NAIL HOLDING SCREW GAMMA3® 8X35MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1320-0090
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "we hereby inform you that yesterday (b)(6) 2022, at 7.45 pm, at the (b)(6) clinic, a gamma nail surgery took place.The procedure was carried out correctly from the early stages of assembly up to the implantation of the distal screw.Everything as usual.When it was time to unscrew the screw that locks the nail to the carbon mask, however, it showed no signs of yielding even by a millimeter.Consequently, the operator had to remove the nail from the patient, hoping that on the operating table it would be easier to remove it from the mask.Not even with the use of a vice were they able to move the screw.At that point the agent, in urgency and to complete the intervention, was forced to take a gamma 3 instrumentation in storage at the aurelia hospital to take it to paideia, where it was sterilized and then, with about two hours of delay from the practice, the operator has finally finished the nail.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "we hereby inform you that yesterday (b)(6) 2022, at 7.45 pm, at the (b)(6), a gamma nail surgery took place.The procedure was carried out correctly from the early stages of assembly up to the implantation of the distal screw.Everything as usual.When it was time to unscrew the screw that locks the nail to the carbon mask, however, it showed no signs of yielding even by a millimeter.Consequently, the operator had to remove the nail from the patient, hoping that on the operating table it would be easier to remove it from the mask.Not even with the use of a vice were they able to move the screw.At that point the agent, in urgency and to complete the intervention, was forced to take a gamma 3 instrumentation in storage at the aurelia hospital to take it to paideia, where it was sterilized and then, with about two hours of delay from the practice, the operator has finally finished the nail.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the inner surface of the returned nail holding screw is damaged possibly would have been caused in an attempt to disassemble.The remnants of dried blood can be observed in the threads indicating the blood percolated in the assembly.The received assembly of devices could not be disassembled as the nail holding screw was found to be stuck.Post decontamination, pre-surgical test was conducted for the returned assembly.Upon inspection, the devices were functionally ok.The nail holding screw could be easily advanced into the target device and the nail.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the issue is most likely user related.The device shows appearance of blood residual on its surfaces, which potentially caused overtightening of the nail in between the interface of the nail and the screw.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAIL HOLDING SCREW GAMMA3® 8X35MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14473871
Report Number0009610622-2022-00201
Device Sequence Number4504223
Product Code LXH
UDI-Device Identifier04546540377661
UDI-Public04546540377661
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1320-0090
Device Catalogue Number13200090
Device Lot NumberKSS217112
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer09/05/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age87 YR
Patient SexMale
-
-