• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER; INTRODUCER, CATHETER Back to Search Results
Model Number 406804
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
During an atrial fibrillation ablation procedure, while inserting dilator into introducer, dilator had punctured the introducer at the sheath tip.A new device was used to continue the procedure with no adverse patient consequences.
 
Additional Manufacturer Narrative
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.The sheath tubing had been stretched and protruded in three locations within the manufactured curve.No punctures were noted.The shapes of the protruded material were consistent with the shape of the dilator distal tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The damage to the sheath is consistent with forcible contact between the dilator and the inner diameter of the sheath during dilator insertion.The cause of the forcible contact remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14475852
Report Number3005334138-2022-00337
Device Sequence Number17327229
Product Code DYB
UDI-Device Identifier05414734203302
UDI-Public05414734203302
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number406804
Device Catalogue Number406804
Device Lot Number7167520
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-