| Model Number |
453564634201 |
| Medical Device Problem Codes |
Defibrillation/Stimulation Problem (1573); Measurement System Incompatibility (2982)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported that the ekg goes flat when using a defibrillator.According to the customer, they replaced the ecg trunk cable and leads without resolve.Customer is requesting onsite support to further evaluate the device.There.
|
| |
|
Event or Problem Description
|
|
The customer reported that the ekg goes flat when using a defibrillator.According to the customer, they replaced the ecg trunk cable and leads without resolve.Customer is requesting onsite support to further evaluate the device.The device was in clinical use at the time the reported issue was discovered.The patient reportedly had to be defibrillated.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips field service engineer (fse) was dispatched to the customer's site to address the reported problem.The device evaluation was performed by the fse on the actual device involved in this complaint.According to service notes, the fse could not reproduce the customer's reported problem.The device was reimaged for customer assurance and tested.The device passed all functional testing and was returned to service.
|
| |
|