| Model Number |
20128E |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that 2 bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: i have two sets of the smartsite extension set filter tubing with ref #(b)(4) that would not prime for staff.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 19oct20223 investigation summary: 2 samples were returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The sets were attached to a bd primary set primed with 85% glycerin / 15% water solution and allowed to prime using gravity.The sets were unable to be primed using gravity.A quality notification was sent to the supplier.During the supplier investigation excess solvent was observed.A quality notification was sent to the manufacturer.The manufacturer investigation found the main root cause to be attributed to double gumming (repeat the solvent application twice).As a preventative action a quality notification was created for this defect.A device history record review for model 20128e lot number 21129190 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on (b)(6) 2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: i have two sets of the smartsite extension set filter tubing with ref #(b)(4) that would not prime for staff.
|
| |
|