ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-65-080-120-P6
Medical Device Problem Codes Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interactions with the heavily calcified and mildly tortuous anatomy resulted in the ratchet being unable to properly engage and release the stent from the delivery system thus resulting in reported activation failure/ deployment failure and the reported mechanical jam.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion in the superficial femoral artery(sfa) with heavy calcification and mild tortuosity.The vessel diameter was 6.5mm and the lesion length was 60mm.Using the cross-over-approach, the vessel was prepped with balloon dilatation at 16 atmospheres (atm) for 30 seconds.A 6.0x120mm supera stent and a non-abbott 6.0x120mm stent were implanted without issue.Then the 6.5x80mm supera self expanding stent system (sess) was advanced to the target lesion and attempted to be deployed; however only 2mm of the stent was able to be released from the sheath.The thumbslide was not able to be advanced.Therefore, the sess was removed under fluoroscopy from the patient.There was no adverse patient effects and no clinically significant delay in the procedure.A non-abbott stent was used to complete the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14478075
Report Number2024168-2022-05524
Device Sequence Number5321290
Product Code NIP
UDI-Device Identifier08717648226267
UDI-Public08717648226267
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberS-65-080-120-P6
Device Catalogue NumberS-65-080-120-P6
Device Lot Number1032961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2022
Initial Report FDA Received Date05/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDING SHEATH: DESTINATION6F; STENT : SUPERA6.0×120; STENT : VIABAHN 6.0×120
Patient SexUnknown
-
-