• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to vyaire medical that the vela ventilator alarmed "xdcr fault (transducer fault), low pip (positive inspiratory pressure) and low ve (ventilation) while on a patient.The patient was transferred to another ventilator.The customer confirmed that there was no patient harm associated with the reported event.
 
Additional Manufacturer Narrative
The customer reported that they will not return the defective unit/part for evaluation.Therefore, no root cause could be determined.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14478803
Report Number2021710-2022-16077
Device Sequence Number4680346
Product Code CBK
UDI-Device Identifier10846446001600
UDI-Public(01)10846446001600(11)20130315
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K093094
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVELA
Device Catalogue Number16530-00
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Initial Report FDA Received Date05/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-