| Model Number |
OTV-S7PROH-HD-12Q |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed due to a faulty charged coupled device.In addition, a failed leak test and punctured cable were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
The customer reported the hd autoclavable camera head display black image.The event occurred during preparation for use and there was no patient harm or user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 12 years since the subject device was manufactured.Based on the results of the investigation, the investigation confirmed no image due to a faulty charge-coupled device.The specific root cause of the faulty charge-coupled device.Could not be determined at this time.Olympus will continue to monitor field performance for this device.
|
| |
|