Model Number 71453183
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Purulent Discharge (1812); Unspecified Infection (1930)
Date of Event 12/10/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tka surgery had been performed on (b)(6) 2016 with a legion porous system, patient felt right knee hematogenous infection.A right knee incision and drainage with revision of articular surfaces treatment was performed to resolve the reported event.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a bilateral articular insert revision with incision and drainage was performed approximately 5.5 years post implantation due to ¿patient felt right knee hematogenous infection¿.The requested clinical documentation has not been provided as of the date of this medical investigation; however, it was reported that the aemb cannot exclude the contribution of the device to the infection.The event occurrence date was reportedly 10-dec.2021 with a status of ¿resolved¿.Without the requested clinical documentation, the root cause of the reported event could not be further assessed.The patient impact beyond the reported infection.I&d and articular insert revision could not be determined.No further medical assessment could be rendered at this time.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that acute post-surgical wound infection have been identified as a possible adverse effect.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN CR HIGH FLEX XLPE SZ 3-4 10MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14480864
Report Number1020279-2022-02562
Device Sequence Number5322520
Product Code JWH
UDI-Device Identifier00885556487341
UDI-Public00885556487341
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71453183
Device Catalogue Number71453183
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date06/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-