• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 8831661001
Medical Device Problem Codes Material Fragmentation (1261); Product Quality Problem (1506); Difficult to Remove (1528); Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, initially there was a problem inserting the cvc (central venous catheter) into the right femoral.It was stated that the catheter immediately had problems and was removed.The following day, the customer reported that the guide wire did not pass.It was stated that it was placed for start of chronic dialysis and placement was said to be ultrasound guided.The functionality was sufficient and with heparin lock.Venous priming was 1.3 and arterial priming was 1.2.The catheter was not repaired.There was no leak and no luer adapter issue.Tego was not utilized.There was no patient symptoms or complications associated with this event.There was no reported patient injury.
 
Event or Problem Description
According to reporter, on (b)(6), 2022 (friday) a contralateral central venous catheter (cvc) was placed and had to be remove the following day for internal organizational problems.During replacement, there was a problem of insertion of the guidewire (includedin the kit).It was stated that the guidewire was pinned (did not pass) inside the catheter and did not come out in the tip of the catheter.When removed the guidewire was folded and all pieces were accounted for.Another catheter of the same batch was implanted and had no problem.It was stated that it was placed for start of acute dialysis and placement was said to be ultrasound guided.The functionality was sufficient and with heparin lock.Venous priming was 1.3 and arterial priming was 1.2.The catheter was not repaired.There was no leak and no luer adapter issue.Tego was not utilized.There was nothing unusual observed on the device prior to use, flushing was done prior to use with no problem, no other defects/damages found on the product, no excessive force was used on the device and only heparin was the cleaning agent used on the device.Intervention/treatment was not required as a result of the event.There was no blood loss, and blood transfusion was not required.There was no patient symptoms or complications associated with this event.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: b3, b5, d9, g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, on (b)(6), 2022 the removal of the catheter was evaluated and to make sure to dialysis the patient, another central venous catheter (cvc) was in placed.The following day (b)(6) 2022, the central venous catheter (cvc) was removed and replaced.During replacement, there was a problem with the guidewire (included in the kit).It was stated that the guidewire was pinned (did not pass) and did not allow replacement.When removed the guidewire was not intact and all pieces were accounted for.It was stated that it was placed for start of chronic dialysis and placement was said to be ultrasound guided.The functionality was sufficient and with heparin lock.Venous priming was 1.3 and arterial priming was 1.2.The catheter was not repaired.There was no leak and no luer adapter issue.Tego was not utilized.There was nothing unusual observed on the device prior to use, flushing was done prior to use with no problem, no other defects/damages found on the product, no excessive force was used on the device, only heparin was the cleaning agent used on the device and intervention/treatment was not required as a result of the event.There was no blood loss, and blood transfusion was not required.There was no patient symptoms or complications associated with this event.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: g1, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the guide wire was unable to be withdrawn and there was a guide wire issue.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS   20101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14481366
Report Number3009211636-2022-00126
Device Sequence Number17327260
Product Code FOZ
UDI-Device Identifier10884521005235
UDI-Public10884521005235
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K943349
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/21/2025
Device Model Number8831661001
Device Catalogue Number8831661001
Device Lot Number2035100043
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2022
Supplement Date Received by Manufacturer05/27/2022
06/17/2022
08/04/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date06/13/2022
06/27/2022
08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-