The customer reported a complaint that "the autopulse platform did not power on intermittently using the autopulse li-ion battery (sn (b)(6))" was not confirmed during the functional testing and archive review.Upon visual inspection, no physical damage was observed.The battery archive was downloaded and reviewed.A review of the battery archive did not show any errors or anomalies related to the reported complaint of not powering up the autopulse platform, however, on (b)(3) 2022, the battery recorded 3 in-out device events.No further events were recorded.The battery was last charged successfully on (b)(3) 2022, and was last used/inserted in the autopulse platform on (b)(3) 2022.During functional testing, the autopulse li-ion battery passed charging in a known good autopulse multi-chemistry battery charger (mcc), and the status leds of the battery showed four green lights after successful charging.The platform was tested with the battery (sn (b)(6)) and the battery performed as intended.
|