ZOLL CIRCULATION AUTOPULSE® LI-ION BATTERY; LI-ION BATTERY, RECHARGEABLE Back to Search Results
Model Number MODEL 100
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the autopulse platform (sn (b)(4)) does not power on with four fully charged autopulse li-ion batteries (sn (b)(4)).The issue was observed intermittently.It is unknown where the problem occurred.However, patient use information was requested but no additional information was provided.Please see the following related mfr reports: mfr 3010617000-2022-00578 for autopulse platform sn (b)(4).Mfr 3010617000-2022-00587 for autopulse li-ion battery sn (b)(4).Mfr 3010617000-2022-00588 for autopulse li-ion battery sn (b)(4).Mfr 3010617000-2022-00589 for autopulse li-ion battery sn (b)(4).
 
Additional Manufacturer Narrative
Zoll has received the autopulse battery for investigation.The battery sn (b)(4) was tested with the returned autopulse platform (sn (b)(4)) and was able to power on the platform without any issue.Further investigation is pending.A follow-up report will be filed when the investigation has been completed.
 
Additional Manufacturer Narrative
The zoll autopulse li-ion battery failure mode and serial number (b)(6) reported in this ccr are covered by capa 22-005.Refer to the capa documentation for details on the failure mode description, technical root cause discussion, and corrective action plan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LI-ION BATTERY
Common Device Name
LI-ION BATTERY, RECHARGEABLE
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key14484388
Report Number3010617000-2022-00590
Device Sequence Number4808429
Product Code MOQ
UDI-Device Identifier00849111001861
UDI-Public00849111001861
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
D163369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0752-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-