| Model Number |
EX3100 |
| Medical Device Problem Codes |
Connection Problem (2900); No Pacing (3268)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented in the clinic for an implant.During the procedure, the pacing system analyzer (psa) unexpectedly disconnected and was unable to provide pacing support.The psa was replaced and the implant procedure was successful.The patient was stable throughout.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of the pacing system analyzer (psa) unexpectedly disconnected and unable to provide pacing support could not be verified.Final analysis found that, the psa was plugged into a testing merlin and powered on successfully.The test performed on the psa revealed that all the resistors (atrium, left ventricle, and right ventricle) were within the impedance levels.No anomalies were found.
|
| |
|