• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X280MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X280MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3425-1280S
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Event or Problem Description
It was reported: "the surgery was almost complete when the aiming device was to be removed, it was not detachable.The aiming device cannot be removed from the long gamma nail.There was a quasi cold connection between the implant and the insertion aid.Consequence: removal of the already fully implanted nail incl.Femoral neck screw and distal locking and re-implantation of a new nail.The procedure was delayed by 1h which lead to higher blood loss.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.Device inspection revealed the following: the returned devices were received in a disassembled state.No visual damage to nail was observed post decontamination, pre-surgical test was conducted for the returned devices.Upon inspection, the devices were functionally ok.While inserting the nail holding screw in the target device slight resistance could be felt but parts could be assembled and disassembled a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the issue is most likely user related.The percolated blood in targeting arm most likely caused cold welding resulting in jamming of the devices making it difficult to disassemble if the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported: "the surgery was almost complete when the aiming device was to be removed, it was not detachable.The aiming device cannot be removed from the long gamma nail.There was a quasi cold connection between the implant and the insertion aid.Consequence: removal of the already fully implanted nail incl.Femoral neck screw and distal locking and re-implantation of a new nail.The procedure was delayed by 1h which lead to higher blood loss.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X280MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14487171
Report Number0009610622-2022-00207
Device Sequence Number11964476
Product Code HSB
UDI-Device Identifier07613252273790
UDI-Public07613252273790
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number3425-1280S
Device Catalogue Number34251280S
Device Lot NumberK08616E
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1