• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL NAIL HOLDING SCREW GAMMA3® 8X35MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL NAIL HOLDING SCREW GAMMA3® 8X35MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1320-0090
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported: "the surgery was almost complete when the aiming device was to be removed, it was not detachable.The aiming device cannot be removed from the long gamma nail.There was a quasi cold connection between the implant and the insertion aid.Consequence: removal of the already fully implanted nail incl.Femoral neck screw and distal locking and re-implantation of a new nail.The procedure was delayed by 1h which lead to higher blood loss.".
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.Device disposition unknown.
 
Event or Problem Description
It was reported: "the surgery was almost complete when the aiming device was to be removed, it was not detachable.The aiming device cannot be removed from the long gamma nail.There was a quasi cold connection between the implant and the insertion aid.Consequence: removal of the already fully implanted nail incl.Femoral neck screw and distal locking and re-implantation of a new nail.The procedure was delayed by 1h which lead to higher blood loss.".
 
Additional Manufacturer Narrative
Please note the correction to d9/h3 as the device was returned for evaluation.Please also note the corrections to h6 method, h6 results and h6 conclusion codes.The reported event could be confirmed.Device inspection revealed the following: the returned devices were received in a disassembled state.The inner surface of the returned nail holding screw is damaged possibly would have been caused in an attempt to disassemble.Fretting marks on the shaft of nail holding screw are visible suggestive of cold welding post decontamination, pre-surgical test was conducted for the returned devices.Upon inspection, the devices were functionally ok.While inserting the nail holding screw in the target device slight resistance could be felt but parts could be assembled and disassembled a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the issue is most likely user related.The percolated blood in targeting arm most likely caused cold welding resulting in jamming of the devices making it difficult to disassemble if the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAIL HOLDING SCREW GAMMA3® 8X35MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14487199
Report Number0009610622-2022-00205
Device Sequence Number20003660
Product Code LXH
UDI-Device Identifier04546540377661
UDI-Public04546540377661
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2019
Device Model Number1320-0090
Device Catalogue Number13200090
Device Lot NumberK03FBAD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-