Model Number 1504-00-107
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 04/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Instability of the prosthesis after post-fracture of the distal femur.Surgical procedure: total knee prosthesis for revision of medial unicompartmental prosthesis.Between the implantation of the cr attune depuy prosthesis and the revision there was a supracondylar periprosthetic fracture from trauma (not connected to the prosthesis or to its instability), which is to be considered the main responsible for the instability that we have come to treat with the revision.For the rest, it was a total explantation of components that were well adhered to the bone.Doi: (b)(6) 2019.Dor: (b)(6) 2022.Unknown knee.
 
Additional Manufacturer Narrative
Product complaint (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM LT SZ 7 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14487592
Report Number1818910-2022-09516
Device Sequence Number14636660
Product Code NJL
UDI-Device Identifier10603295041085
UDI-Public10603295041085
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1504-00-107
Device Catalogue Number150400107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 7 6MM.; ATTUNE CR FEM LT SZ 7 CEM.; ATTUNE FB TIB BASE SZ 7 CEM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexUnknown
-
-