| Model Number |
03.033.004 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, the surgeon was performing a piriformis femoral recon nail.After finishing the opening drill bit using cannulated drill bit flexible was broken near the proximal end of the drill bit.Later in the case, the retention pin screwdriver was used and it was noticed that the tip of the retention was missing.It did not break in this case and it was unsure when it was broken.The case was successfully completed without any surgical delay.No harm caused to patient and no broken fragments in patient.This report is for one (1) 14mm cannulated drill bit flexible.This is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary = > the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Photos of the item was received as well and the findings were consistent with physical device investigation.Visual analysis of the returned sample revealed that 14mm cannulated drill bit flexible was observed broken at the flexible shaft near the coupling.No other defect was found.A dimensional inspection for the 4mm cannulated drill bit flexible was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was [confirmed as the observed condition of the 4mm cannulated drill bit flexible would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review = > a manufacturing record evaluation was performed for the part: 03.033.004-us lot: 5l56857; and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|