| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that they were using this device during a patient event and when they went to charge the device to deliver defibrillation energy, the device took too long to charge.The user(s) switched to a back up device and used that to continue patient care.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer informed stryker that no further patient information is available.Patient fields in which information was not provided were intentionally left blank.The device was not returned to stryker for evaluation.The cause of the reported issue could not be determined.
|
| |
|