• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FUJIFILM HEALTHCARE CORPORATION OASIS XP 1.2T OPEN MRI SYSTEM; MAGNETIC RESONANCE DIAGNOSTIC DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FUJIFILM HEALTHCARE CORPORATION OASIS XP 1.2T OPEN MRI SYSTEM; MAGNETIC RESONANCE DIAGNOSTIC DEVICE Back to Search Results
Model Number N/A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Tinnitus (2103)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
On april 28, 2022, fujifilm healthcare corporation received a complaint regarding the oasis xp 1.2t open mri system.It was reported that a female patient complained of ringing in the left ear following a lower extremity scan.During the scan, the patient was wearing headphones and listening to jazz music.The technician confirmed that the patient could hear the music at the time of the scan.The patient stated that she did not notice the ringing until she was home.No hearing loss was reported and no further information was provided.There is no death or serious injury associated with the event.As such, this report is being submitted in an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OASIS XP 1.2T OPEN MRI SYSTEM
Common Device Name
MAGNETIC RESONANCE DIAGNOSTIC DEVICE
Manufacturer (Section D)
FUJIFILM HEALTHCARE CORPORATION
2-1 shintoyofuta
kashiwa-shi, chiba-ken 277-0 804
JA  277-0804
MDR Report Key14489913
Report Number1528028-2022-00038
Device Sequence Number20017536
Product Code LNH
UDI-Device Identifier04580292768186
UDI-Public(01)04580292768186(11)191015(250)G210090219
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/28/2022
Device Age31 MO
Event Location Outpatient Diagnostic Facility
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/24/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-