• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-4-12-HX-SS
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An axium prime coil was returned for analysis within a shipping box and within a sealed biohazard pouch.Visual inspection/damage location details: the non-medtronic micro catheter (sl-10 stryker) used in the event was not returned.The inner diameter (id) of the sl 10 microcatheter was found to be 0.0165¿ (per an online source) per axium prime coil instructions for use (ifu): ¿axium prime detachable coils should be delivered through a microcatheter with a minimum inside diameter of 0.0165¿-0.017¿ with two marker bands.¿.Therefore, the sl 10 micro catheter was found to be compatible for use with the axium prime coil.The axium prime coil was returned without introducer sheath.The actuator interface was found to be intact but kinked.The axium prime coil pushwire was found to be bent at ~3.6cm and broken at ~143.5cm from proximal end.The implant coil was found to be damaged and not returned.No other anomalies were observed.Testing/analysis (including sem reports): the axium prime coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck in catheter¿ could not be confirmed as the axium prime coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.Possible contributing factors to ¿coil resistance/stuck in catheter¿ include severely tortuous patient anatomy, failure to hydrate the axium implant coil prior to use, and lack of continuous flush during delivery.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the coil was stuck in the proximal section of the catheter.The implant coil pushed smoothly out from the introducer sheath.There was no coil damage.No patient symptoms or further complications were reported as a result of this event.Ancillary devices include a navien072 guide catheter, sl-10 microcatheter, and synchro14 guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14489928
Report Number2029214-2022-00880
Device Sequence Number4736294
Product Code KRD
UDI-Device Identifier00847536032323
UDI-Public00847536032323
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2024
Device Model NumberAPB-4-12-HX-SS
Device Catalogue NumberAPB-4-12-HX-SS
Device Lot NumberB260237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Initial Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-