• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24628
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2022
Event Type  malfunction  
Event Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous left forearm cephalic vein.A 5.00mm/2.0cm/50cm peripheral cutting balloon was selected for use.During the procedure, first inflation was performed at 10atm for 60 seconds, however the balloon ruptured upon deflation.The device was simply pulled out from the patient's body.The procedure was completed with another of same device.No complications reported and patient was good post procedure.
 
Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The rated burst pressure for this device is 10 atmospheres as per pcb2cm specification.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 8mm distal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination observed no damage to the tip or blades.All other blades were present and fully bonded to the balloon surface.A visual and tactile examination identified no kinks or damage to the shaft of the device.This concludes the product analysis.
 
Event Description
It was reported that a balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous left forearm cephalic vein.A 5.00mm/2.0cm/50cm peripheral cutting balloon was selected for use.During the procedure, first inflation was performed at 10atm for 60 seconds, however the balloon ruptured upon deflation.The device was simply pulled out from the patient's body.The procedure was completed with another of same device.No complications reported and patient was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14490453
MDR Text Key292569214
Report Number2134265-2022-05877
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number24628
Device Catalogue Number24628
Device Lot Number0028116893
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-