Model Number 24691
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter city: (b)(6).
 
Event Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified superficial femoral artery.A 3mm x 40mm x 146cm coyote es balloon catheter was advanced for dilation.The balloon was inflated twice at 10 atmospheres (atm) for 10 seconds and 14 atm for 5 seconds.However, during second inflation at 14 atm for 5 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications reported and the patient condition was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE ES
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14490488
MDR Text Key292569072
Report Number2134265-2022-05892
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08714729767251
UDI-Public08714729767251
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K093636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24691
Device Catalogue Number24691
Device Lot Number0027129295
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-