MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK AXS CATALYST 7 .068IN ID X 132CM - US; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK AXS CATALYST 7 .068IN ID X 132CM - US; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number IC068132
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual/microscopic inspection: the catheter shaft was seen to be broken from near the strain relief.The catheter shaft was seen to be kinked 41cm from the hub.Functional inspection: a 0.067" pin gauge was verified to meet the catheter shaft when inserted into the catheter hub.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported hub break was not confirmed during the analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.As per the additional information, the device was prepared as per the dfu and there were no anomalies noted to the device during initial inspection and preparation and continuous flush was maintained for the duration of the procedure.During the analysis, catheter shaft was seen to be broken near the strain relief and catheter shaft was seen to be kinked.It is probable that the device was damaged during the clinical procedure.An assignable cause of procedural factors will be assigned to the as analyzed codes, 'catheter shaft broken/fractured during use' and 'catheter shaft kinked/bent' as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.An assignable cause of 'not confirmed' was assigned to the as reported defect 'catheter hub broken/fractured' as the issue included a returned product review (visual, physical, and/or performance testing) which showed no evidence of either the alleged issue(s) or any defect which could have contributed to the event.
 
Event or Problem Description
The subject aspiration catheter was returned for analysis and it was discovered that the subject catheter shaft was broken/fractured during use.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS CATALYST 7 .068IN ID X 132CM - US
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14490655
Report Number3008881809-2022-00242
Device Sequence Number17473192
Product Code NRY
UDI-Device Identifier07613327417227
UDI-Public07613327417227
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K183464
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/12/2023
Device Model NumberIC068132
Device Catalogue NumberIC068132
Device Lot Number23140450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Initial Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYNCHRO GUIDEWIRE (STRYKER)
Patient SexUnknown
-
-