• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 BROACH CORAIL AMT 10; HIP INSTRUMENTS : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 BROACH CORAIL AMT 10; HIP INSTRUMENTS : BROACHES Back to Search Results
Model Number L20410
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
Unable to assemble broaches with trial necks.Surgical delay of 5 minutes occurred.Another instrument was used to finish the case.No adverse affects on patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was returned for investigation.After visual examination of the piece, it was identified that the post of the device was scratched.A functional test was performed and the device was able to be assembled with the mating neck trail.The reported condition was not confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROACH CORAIL AMT 10
Common Device Name
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR   69801
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14491889
Report Number1818910-2022-09574
Device Sequence Number12231053
Product Code HTQ
UDI-Device Identifier10603295258124
UDI-Public10603295258124
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL20410
Device Catalogue NumberL20410
Device Lot NumberSO2024109
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer06/09/2022
06/23/2022
08/02/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date06/12/2022
06/24/2022
08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BROACH CORAIL AMT 10.; BROACH CORAIL AMT 11.; BROACH CORAIL AMT 13.
Patient SexUnknown
-
-