| Model Number |
L20410 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Unable to assemble broaches with trial necks.Surgical delay of 5 minutes occurred.Another instrument was used to finish the case.No adverse affects on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: the device associated with this report was returned for investigation.After visual examination of the piece, it was identified that the post of the device was scratched.A functional test was performed and the device was able to be assembled with the mating neck trail.The reported condition was not confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|
Search Alerts/Recalls
|