| Model Number |
07P7420 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely elevated ldh results generated on the alinity c processing module for one patient.The initial result was 731 u/l, repeated 215 and 202 u/l (normal range 125 to 220 u/l).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the suspect medical device was changed from alinity c lactate dehydrogenase reagent kit, list number 07p74-20, and manufacturing site wiesbaden to alinity c processing module, list number 03r67-01, and manufacturing site of irving.Mdr number 3016438761-2022-00262-00 has been submitted and all further information will be documented under that mdr number.
|
| |
|