Model Number 39200
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent premature deployment occurred.The target lesion was located in iliac artery.A 8x40x120 epic vascular stent was advanced for treatment.However, during the procedure, when the stent was flushed, the tip of the stent automatically deployed.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 6mm from the distal end of the middle sheath.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found no damage that could have contributed to the premature deployment.
 
Event or Problem Description
It was reported that stent premature deployment occurred.The target lesion was located in iliac artery.A 8x40x120 epic vascular stent was advanced for treatment.However, during the procedure, when the stent was flushed, the tip of the stent automatically deployed.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14492828
Report Number2134265-2022-05650
Device Sequence Number6150795
Product Code NIO
UDI-Device Identifier08714729805649
UDI-Public08714729805649
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0028373239
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date05/24/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-